trace SN cube

Elemental analyzer trace SN cube for the analysis of trace concentrations of nitrogen and sulfur in petroleum products

trace SN cube

The continuously increasing requirements for environmental protection mean further decreasing threshold values for sulfur or nitrogen in fuels. Instruments for measuring these components have to correspond to these future requirements in their performance. Therefore, Elementar as the technology leader in elemental analysis, has developed the top level instrument for sulfur and nitrogen analysis which safely and accurately analyzes sulfur or nitrogen concentrations in mineral oils and other petro- chemical products in the lower ppb range and exceeds the requirements of today’s standards by far.

Device complies with many modern international and Russian standards

S

N

ГОСТ Р ЕН ИСО 20846-2006 "Нефтепродукты. Методы определения содержания серы методом ультрафиолетовой флуоресценции"

ASTM D 4629 Standard Test Method for Trace Nitrogen in Liquid Petroleum Hydrocarbons by Syringe/Inlet Oxidative Combustion and Chemiluminescence Detection

ASTM D 5453 Standard Method for Determination of Total Sulfur in Light Hydrocarbons, Motor Fuels and Oils by Ultraviolet Fluorescence

ASTM D 7184 Standard Test Method for Ultra Low Nitrogen in Aromatic Hydrocarbons by Oxidative Combustion and Reduced Pressure Chemiluminescence Detection

ASTM D 6667 Standard Test Method for Determination of Total Volatile Sulfur in Gaseous Hydrocarbons and Liquefied Petroleum Gases by Ultraviolet Fluorescence

ASTM D 6069 Standard Test Method for Trace Nitrogen in Aromatic Hydrocarbons by Oxidative Combustion and Reduced Pressure Chemiluminescence Detection

ASTM D 7551 Total Volatile Sulfur in Gaseous Hydrocarbons and Liquefied Petroleum Gases and Natural Gas by Ultraviolet Fluorescence

UOP 981 Trace Nitrogen in Liquid Hydrocarbons by Oxidative Combustion with Chemiluminescence Detection

ASTM D 7183 Standard Test Method for Determination of Total Sulfur in Aromatic Hydrocarbons and Related Chemicals by Ultraviolet

DIN 51444 Testing of petroleum products - Determination of nitrogen - Oxidative combustion method with chemiluminescence detector

DIN EN 15486 Ethanol as a blending component for petrol - Determination of sulfur content - Ultraviolet fluorescence method

 

DIN ISO 20846 Petroleum products - Determination of sulfur content of automotive fuels - Ultraviolet fluorescence method

 

UOP 987 Low Trace Sulfur in Liquid Hydrocarbons by Oxidative Combustion with Ultraviolet Fluorescence Detection

 

The method of determining is a direct the dosing of liquid or gaseous samples by using the catalytic combustion at high temperature up to 1150 ° C and subsequent determination of sulfur dioxide by using UV fluorescence detector,  the determination of nitric oxide by using chemiluminescent detector (CLD).

Analysis modes:

  • determination of sulfur
  • determination of nitrogen
  • joint determination of sulfur and nitrogen
  • joint determination chlorine* and sulfur (or nitrogen)

* Determination of chlorine by electrochemical detector.

Type of samples:

  • liquid
  • gaseous
  • Liquefied Gases
  • Solid

Sample feeding:

  • manual input of liquid samples using a microsyringe
  • manual input of the gaseous sample through a syringe
  • manual input of solid samples
  • The liquid autosampler Vario Liquid Sampler (VLS), does not require operator intervention and can be fitted with a thermostat. Autosampler VLS can hold up to 50 standard vials of 1.5 ml. The feed rate of the sample into the tube combustion:: 40 µl/sec.
  • The company Elementar offers Gas Module for compressed gases, for liquefied gases - LPG Module.
    Trace SN Cube Autosampler
  • Solid sampler  (80 or 120 samples). Maximum sample weight up to 1 g.

The maximum sample volumes up to 100 µl of liquid samples, or 10 ml for sample gas. Time of dosing only a few seconds at full combustion provides a particularly sharp peaks.

Measuring range:

S

N

Cl

0,1 – 200 ppm (syringe 100 µl)

2 – 2000 ppm (syringe 10 µl)

0,1 – 200 ppm (syringe 100 µl)

2 – 2000 ppm (syringe 10 µl)

0,05 – 50 µg

Detection limit (LOD):

  • 6 ppb for sulfur in liquid samples
  • 100 ppb for sulfur in gaseous samples
  • 12 ppb for nitrogen

Technology«Purge-and-Trap» (optional): To improve the reproducibility of the results, as well as traces of sulfur analysis on the background of a large amount of nitrogen can be installed adsorption column for sulfur dioxide to separate it from a peak of nitrogen dioxide.

S peak without using the adsorption column S peak using adsorption column
S peak without using the adsorption column S peak using adsorption column

Precision: < 2 %.

Management and processing of results: via PC.

Necessary gases:

  • for the determination of sulfur: synthetic air, free of SO2 and CO2
  • for the determination of nitrogen: a mixture of oxygen (20%) and argon (80%), without nitrogen impurities.